Specifications
| Animal Species | Rat, mouse, monkey, dog, rabbit, guinea pig, cat, horse |
|---|---|
| Capillary Mode | CBC+5 Diff(WBC)+Ret* with 40uL sample |
| Manual Gating Function | Manual gating of cell populations by parameter of RBC, PLT, WBC, DIFF and RET* is available. |
| Data Storage | Test data: 3,000 samples Test information: 1,000 information QC files: 16 files Data export by CSV format is available |
| Part 11software | FDA 21 CFR Part 11compliant software is prepared as option. |
*: XT-2000iV only
XT-2000iV/1800iV is Animal Blood Cell Analysis Software for XT-2000i/XT-1800i
XT-2000i/1800i main unit specification
| Parameters | WBC,RBC,HGB,HCT,MCV,MCHC,PLT,NEUT%,LYMPH%, MONO%,EO%,BASO%,NEUT#,LYMPH#,MONO#,EO#, RET%*,RET#* ,RDW-SD,RDW-CV,PDW,MPV,PCT,P-LCR HFR*,MFR*,LFR*,IRF* |
|---|---|
| Throughput | 80 samples/hour |
| Aspiration volume | Manual open mode: 85µL CP sampler mode: 150µL Capillary mode: 40µL |
*: XT-2000iV only
PDW,PCT,P-LCR,HFR,MFR and LFR are not available in USA.
XT-2000i/1800i Dimensions/Weights
| Dimensions W x H x D |
Weights Kg |
Power source (50Hz/60Hz) |
Consumption (50Hz/60Hz) |
||
|---|---|---|---|---|---|
| Main unit | 530 x 630 x 500 | 52 | AC100-240V | 250VA | |
| Sampler | 520 x 110 x 220 | 7 | |||
| Pneumatic unit | 280 x 400 x 355 | 17 | AC100-117V | 50Hz 230VA | |
| 60Hz 280VA | |||||
| AC220-240V | 50Hz 220VA | ||||
| 60Hz 250VA | |||||